The 4WFBS120 is a 4-wire, full-bridge module that provides completion resistors for a strain gage or other sensor that acts as a single variable resistor. This module should be used to measure sensors that have a 120 ohm nominal resistance.
Read MoreThe 4WFBS120 completes a full Wheatstone bridge. It can also be used to complete the back half of a Wheatstone bridge for use in a ¼ bridge strain circuit (one active element) using a dummy gage, or in a ½ bridge strain circuit (two active elements). The Wheatstone bridge circuit converts small changes in resistance to an output voltage that our dataloggers can measure.
The 4WFBS120 includes two external pins that allow customers to perform shunt calibrations, which are used to correct for sensitivity errors.
Used With | 4-wire strain gages or other full bridge measurements that have a 120 Ω nominal resistance |
Dimensions | 1.5 x 1.5 x 2.7 cm (0.6 x 0.6 x 1.0 in.) for body with prongs |
Weight | 6 g (0.2 oz) |
2:1 Resistive Divider |
|
Resistance | 1 kΩ/1 kΩ |
Tolerance | Ratio: ±0.01% (@ 25°C) |
Maximum Temperature Coefficient | Ratio: 0.5 ppm/°C (-55° to +85°C) |
Power Rating | 0.1 W (@ 70°C) per element |
Completion Resistor (bulk metal foil) |
|
Resistance | 120 Ω |
Tolerance | ±0.01% (@ 25°C) |
Maximum Temperature Coefficient | 0.8 ppm/°C |
Power Rating | 0.25 W (@ 70°C) |
Please note: The following shows notable compatibility information. It is not a comprehensive list of all compatible products.
Product | Compatible | Note |
---|---|---|
21X (retired) | ||
CR10 (retired) | ||
CR1000 (retired) | ||
CR10X (retired) | ||
CR200X (retired) | ||
CR211X (retired) | ||
CR216X (retired) | ||
CR23X (retired) | ||
CR300 | ||
CR3000 (retired) | ||
CR310 | ||
CR350 | ||
CR500 (retired) | ||
CR5000 (retired) | ||
CR510 (retired) | ||
CR6 | ||
CR800 (retired) | ||
CR800 (retired) | ||
CR800 (retired) | ||
CR850 (retired) | ||
CR850 (retired) | ||
CR850 (retired) | ||
CR850 (retired) |
Product | Compatible | Note |
---|---|---|
CR9050 (retired) |
The 4WFBS120 is compatible with our multiplexers.
The 4WFBS120 uses one differential analog input; an adjacent analog ground channel accepts the "ground prong" of the 4WFBS120. Not all data logger terminal strips have this sequence on all channels, so please check your data logger's wiring panel to confirm channel assignments, especially if you plan to use multiple Terminal Input Modules.
The lead wire that emanates from the head of the 4WFBS120 connects to a data logger excitation channel.
Please note that if a CR10X will be used to measure thermocouples in the same application, newer CR10X Wiring Panels and the CR10XTCR Reference Thermistor (with its accompanying Thermocouple Cover) are deep enough to cover the data logger input channels with a Terminal Input Module attached; older CR10TCRs are not deep enough.
The 4WFBS120 is measured with the BrFull Instruction in CRBasic, and with Instruction 6 (Full Bridge) in Edlog.
Number of FAQs related to 4WFBS120: 1
The ST350 is designed as a full-bridge strain transducer with four active elements that can be directly connected to a datalogger (or a multiplexer connected to a datalogger) without adding a completion module such as the 4WFBS350 or 4WFBS120.