Services Available | |
---|---|
Repair | No |
Calibration | No |
Free Support | No |
The Q7.1 measured the difference between incoming and outgoing radiation. A thermopile output a voltage proportional to the temperature difference between the top and bottom surfaces of the radiometer.
Contact REBS (manufacturer of the Q7.1) for repair and recalibration: 206-624-7221
Read MoreThe Q7.1 measured the difference between incoming and outgoing radiation. A thermopile output a voltage proportional to the temperature difference between the top and bottom surfaces of the radiometer.
As shading and reflection represent sources of error to the net radiation measurement, the Q7.1 was typically deployed at some distance to the primary instrument mount. A simple vertical pipe affixed with a CM210 mount was recommended for this purpose.
Nominal calibration factors: 9.6 W m-2 mV-1 for positive values; 11.9 W m-2 mV-1 for negative values.
Please note: The following shows notable compatibility information. It is not a comprehensive list of all compatible products.
Product | Compatible | Note |
---|---|---|
21X (retired) | ||
CR10 (retired) | ||
CR1000 (retired) | ||
CR10X (retired) | ||
CR200X (retired) | ||
CR211X (retired) | ||
CR216X (retired) | ||
CR23X (retired) | ||
CR3000 (retired) | ||
CR500 (retired) | ||
CR5000 (retired) | ||
CR510 (retired) | ||
CR800 (retired) | ||
CR800 (retired) | ||
CR850 (retired) | ||
CR850 (retired) | ||
CR850 (retired) | ||
CR850 (retired) | ||
CR9000 (retired) | ||
CR9000X (retired) |
Number of FAQs related to Q7.1-L: 2
The information included on a calibration sheet differs with each sensor. For some sensors, the sheet contains coefficients necessary to program a datalogger. For other sensors, the calibration sheet is a pass/fail report.
This depends on the information contained in the calibration sheet: