The CS10 monitors ac current along an electrical wire for the range of 0.15 A to 200 A. It outputs a millivolt signal that many of our dataloggers can directly measure.
Read MoreThe CS10 uses CR Magnetic's CR8459 Current Transformer to detect the ac current along an electrical wire using the magnetic field that is generated by that current. The CS10 is external to the wire jacket and has no direct electrical connection to the system.
The CS10 is recommended for measurements that do not require high accuracy, such as motor or generator load condition monitoring, efficiency studies, intermittent fault detection, and rough submetering.
Measurement Range | 0.15 to 200 A |
Frequency | 50 and 60 Hz |
Insulation Resistance | 100 MΩ (@ 500 Vdc) |
High Potential | 2000 V |
Rated Current | 200 A |
Accuracy with 10 Ω Maximum Burden (resistive) | Typically ±5% of actual value (with provided multiplier) |
Operating Temperature Range | -25° to +55°C |
Case Material | Polypropylene resin |
Construction | Epoxy encapsulated |
Outer Diameter | 4.8 cm (1.89 in.) |
Inner Diameter | 1.9 cm (0.75 in.) |
Height | 1.7 cm (0.67 in.) |
Please note: The following shows notable compatibility information. It is not a comprehensive list of all compatible products.
Product | Compatible | Note |
---|---|---|
21X (retired) | ||
CR10 (retired) | ||
CR1000 (retired) | ||
CR10X (retired) | ||
CR200X (retired) | ||
CR211X (retired) | ||
CR216X (retired) | ||
CR23X (retired) | ||
CR3000 (retired) | ||
CR5000 (retired) | ||
CR5000 (retired) | ||
CR510 (retired) | ||
CR800 (retired) | ||
CR800 (retired) | ||
CR800 (retired) | ||
CR800 (retired) | ||
CR850 (retired) | ||
CR850 (retired) | ||
CR850 (retired) | ||
CR850 (retired) | ||
CR9000 (retired) | ||
CR9000X (retired) |
Number of FAQs related to CS10-L: 4
Expand AllCollapse All
Most Campbell Scientific sensors are available as an –L, which indicates a user-specified cable length. If a sensor is listed as an –LX model (where “X” is some other character), that sensor’s cable has a user-specified length, but it terminates with a specific connector for a unique system:
If a sensor does not have an –L or other –LX designation after the main model number, the sensor has a set cable length. The cable length is listed at the end of the Description field on the product’s Ordering tab. For example, the 034B-ET model has a description of “Met One Wind Set for ET Station, 67 inch Cable.” Products with a set cable length terminate, as a default, with pigtails.
If a cable terminates with a special connector for a unique system, the end of the model number designates which system. For example, the 034B-ET model designates the sensor as a 034B for an ET107 system.
Many Campbell Scientific sensors are available with different cable termination options. These options include the following:
Note: The availability of cable termination options varies by sensor. For example, sensors may have none, two, or several options to choose from. If a desired option is not listed for a specific sensor, contact an application engineer at Campbell Scientific for assistance.