Analog Input Peripherals / Granite Volt 108
Granite Volt 108 8- or 16-Channel 5V Analog Input Module
Ideal for Expansion of Analog Measurements
Interfaces directly to a PC for quick setup and data acquisition
weather applications supported water applications supported energy applications supported gas flux & turbulence applications supported infrastructure applications supported soil applications supported

Overview

The Granite™ Volt 108 easily expands the channel count on your data-acquisition system. It has eight differential or 16 single-ended input channels, as well as two excitation channels. The Volt 108 provides programmable, precision excitation for standard sensor measurements and power outputs for controlling sensors and peripherals. This module features a 24-bit, analog-to-digital converter with low-noise performance for superior analog measurements.

Benefits and Features

  • 24-bit ADC and low-noise inputs for increased measurement quality
  • Distributed data acquisition
  • Channel count expansion via the CPI bus on Campbell Scientific data loggers
  • Easier to program than traditional multiplexers
  • Programmable noise rejection
  • CANbus 2.0 A/B output available with the Extended Duty (-XD) version
  • USB 2.0 interface for PC-based operation with Campbell Scientific Surveyor software

Images


3D/CAD Files:

Specifications

Power Requirements 9.6 to 32 Vdc voltage
Mounting Standard 1-in. grid (DIN rail mounting available)
Accuracy
  • ±(0.04% of reading + offset) 0° to 40°C
  • ±(0.06% of reading + offset) -40° to +70°C
  • ±(0.08% of reading + offset) -55° to +85°C
Number of Channels 8 differential or 16 single-ended
Operating Temperature Range
  • -40° to +70°C (standard)
  • -55° to +85°C (extended)
Maximum Scan Rate
  • 1 channel at 1 kHz
  • 8 channels at 333 Hz
Input Range ±5000 mV, ±1000 mV, and ±200 mV
Period Averaging Traditional period averaging on analog input channels
CPI For data logger connection. Baud rate selectable from 50 kbps to 1 Mbps. (Allowable cable length varies depending on baud rate, number of nodes, cable quality, and noise environment, but can be as long as 700 m under proper conditions.)
USB USB 2.0 full speed connection available for attaching to a PC. (Port is used to configure the module and download updates via our Device Configuration Utility.)
Warranty One year against defects in materials and workmanship
Dimensions 20.3 x 12.7 x 5.1 cm (8 x 5 x 2 in.)
Weight 0.8 kg (1.75 lb)

Typical Current Drain

Sleep <1 mA
Active 1 Hz Scan 2 mA (estimated)
Assumes one single-ended measurement with the first notch frequency (fN1) at 30 kHz
Note: Any sensor excitation or switched power loads will be additive to this value.
Active 20 Hz Scan 20 mA
Assumes one single-ended measurement with the first notch frequency (fN1) at 30 kHz
Note: Any sensor excitation or switched power loads will be additive to this value.
Active 1 kHz Scan 67 mA
Note: Any sensor excitation or switched power loads will be additive to this value.

Voltage/Current Excitation Outputs

Voltage Excitation ±5 V (@ 50 mA)
Current Excitation ±2.5 mA (±5 V compliance voltage)
Number of Voltage/Current Excitation Outputs 2

General Purpose Outputs

Number of SW5V Outputs 2
SW5V Output Resistance 30 Ω
Number of SW12V Outputs 1
Typical Limit of SW12V Outputs 200 mA
Minimum Limit of SW12V Outputs 180 mA
Number of 12V Outputs 1
Typical Limit of 12V Outputs 200 mA
Minimum Limit of 12V Outputs 180 mA

Compatibility

Please note: The following shows notable compatibility information. It is not a comprehensive list of all compatible products.

Dataloggers

Product Compatible Note
CR1000X Requires OS version 4 or greater.
CR6 Requires OS version 10 or greater.

Miscellaneous

Product Compatible Note
Granite 10 Requires OS version 1 or greater.
Granite 6 Requires OS version 1 or greater.
Granite 9 Requires OS version 1 or greater.

Additional Compatibility Information

The Volt 108 is compatible with CPI-enabled data loggers only.


Related FAQs

Number of FAQs related to Granite Volt 108: 8

Expand AllCollapse All

  1. in Burst Mode operation, repeated measurements are performed on the same channel for X number of times, then there is a waiting period for the measurements to be processed and transmitted to the data logger. No measurements can be made during this processing period. The maximum capture rate is 30 kHz on a single channel for a duration  of 2.18 seconds (absolute maximum capability).

  2. Yes, both 11-bit and the 29-bit (extended) message identifiers are supported.

  3. Yes, a specific CAN ID can be selected for each module.

  4. in continuous scan mode (normal operation), 1 kHz on 1 channel is typical.

  5. The CPI bus speed is adjustable in your CRBasic data logger program. Use the CPISpeed() instruction in your CRBasic program to adjust the CPI bus bandwidth to meet the following maximum combined (total) Ethernet cable lengths:

    • 1000 kB/s for maximum combined Ethernet cable lengths of 15.2 m (50.0 ft)
    • 500 kB/s for maximum combined Ethernet cable lengths of 61 m (200 ft)
    • 250 kB/s for maximum combined Ethernet cable lengths of 152.4 m (500 ft)
  6. The Volt 108/116 XD and Temp 120 XD output CAN2.0b data frames in Big Endian byte format.

  7. All outputs except strain gauge calculated microstrain are supported in CANOut:

    • Volt SE/Diff output in mV
    • BridgeFull in mv/V
    • BrHalf3W and BrHalf4W in Rs/Rf
    • Temp 120 output in desired temperature units

    Note: Strain values are in mV/V and either need a data logger or post processing to calculate microstrain.

  8. Only the XD versions (not the SD) support CANOut 2.0b. To enable this function, use our SURVEYOR software. When SURVEYOR recognizes an XD module, a CAN Enable menu appears with a “CAN Enabled” slider switch.

Listed Under

Other Accessory for the following products: